Dionex RFIC-ER 淋洗液再生啟動套件
使用的Thermo Scientific戴安RFIC-ER淋洗液再生啟動套件為戴安ICS-1100和ICS-1600包含所有為6至12個月經營戴安RFIC-ER系統不間斷所需的耗材和非消耗性物品。選擇套件包括油管,控制器和軟件的Dionex ICS-1000,-1500,-2000或-3000系統升級到RFIC-ER系統中,除了一組完整的消耗品的需要。
Dionex RFIC-ER 淋洗液再生啟動套件
訂貨信息:
RFIC-ER | |
Eluent Regeneration Startup Kit, Anion, for ICS-1100 and ICS-1600 | 069570 |
Eluent Regeneration Startup Kit, Cation, for ICS-1100 and ICS-1600 | 069569 |
RFIC-ER Anion Startup Kit | 067797 |
RFIC-ER Cation Startup Kit | 067798 |
配件 | |
RFIC-ER Anion Consumables Kit | 067791 |
RFIC-ER Cation Consumables Kit | 067792 |
Allows Non-Stop Operation of an IC System for Up to Four Weeks at a Time
RFIC-ER is a low-cost alternative to RFIC-EG systems. Rather than generate eluent electrolytically using deionized water, RFIC-ER uses the electrolytic suppressor to regenerate the starting eluent. Eluent is prepared in the normal manner. After use and regeneration, the eluent is passed through a series of eluent purifier columns to produce pure eluent. The purified eluent is returned to the eluent bottle. One batch of original eluent can provide up to four weeks of non-stop operation.
Always on, always ready capability for ICS-1100, -1600, -2100, and -5000systems
System remains fully equilibrated and calibrated for long periods of time
Simple plumbing and easy-to-understand schematic
Low cost of ownership
Ideal for the analysis of common anions and cations in drinking, surface and ground water samples
Operator errors are significantly reduced or eliminated
Eluent Regeneration System Specifications | |
---|---|
Eluent Controller | |
Dimensions (h × w × d): | 13.6 in × 4.1 in × 4 in 34.5 cm × 10.5 cm × 10 cm |
Weight: | 1.76 lb 0.8 kg |
Maximum Operating Pressure: | 3000 psi post pump 200 psi post suppressor |
Back Pressure: | Less than 335 psi at 1.0 mL/min |
Maximum Flow Rate: | 2.0 mL/min |
Eluent Compatibility: | Carbonate eluents up to 20 mM MSA eluents up to 34 mM |
Solvent Compatibility: | Not compatible with organic solvents in mobile phase |
Temperature Range: | 5 °C to 40 °C |
Eluent Purification Column | |
Lifetime: | 91 days using AS23 chemistry |
Analyte Trap Columns | |
Capacity: | 130,000 μL (e.g., 5200 × 25 μL injections) |