當前位置:北京恒遠安諾科技有限公司>>電工校準儀器>>其他校準儀器>> Model 828電壓電流校正器
FEATURES
AC/DC Output
Wide Range - 1 Millivolt to 1100 Volts,
1 Microampere to 5 Amperes
High Accuracy
Ease of Operation
High Stability
Direct Reading Percent Deviation of
Output Control
High Burden and Compliance Ratings
Short Circuit and Overburden Protection
Continuous Resolution Run-up for Analog Meter Friction Testing
Compact Size for Ease of Portability and Minimum Bench Space
Requirements
Rapid Warm-up
APPLICATIONS
Calibration of: | Designed for use in: |
Analog Panel Meters | Production Line Testing |
Portable Volt-Amp Multimeters | Laboratory Calibration |
3 1/2 Digit Panel Meters | Field Calibration Service |
Transducers | Plant Calibration |
Recorders |
DESCRIPTION
SPECIFICATIONS
OUTPUT SPECIFICATIONS
Full Scale Ranges: | Four: 1, 10, 100, and 1000 V. |
Overrange Limits: | 1100 volts on 1000 V range; 200% of full-scale value on all other ranges. |
Stability: | ?.05%/8 hours; within rated accuracy for 6 months. |
Voltage Burden: | 5 VA on 1 V range; 10 VA on all other ranges. |
Full Scale Ranges: | Five: 1, 10, and 100 mA; 1 and 10A |
Overrange Limits: | 5 amps on 10A range; 200% of full-scale value on all other ranges. |
Compliance: | 1 volt on 10 A range; 5 volts on all other ranges. |
ACCURACY
Dc Voltage | ?0.05% of setting + 0.05% of range) |
Ac Voltage | ?0.05% of setting + 0. 1 % of range) |
Current | ?0. 1 % of setting + 0. 1 % of range) |
GENERAL
Total Harmonic Distortion | Less than 0.3% at full-scale. |
Ripple (Dc Operation Only) | Less than 1 % rms at full-scale with 1 MW load. |
Frequencies: | 50, 60, and 400 Hz. |
Accuracy: | 50 and 400 Hz: 眑 % 60 Hz: ?.5% |
Frequency Limits: | 50 to 440 Hz. |
Input Voltage | 3 volts maximum |
Range: | ?% of setting, adjustable via front panel control. |
Accuracy: | 眑 /2 division (0.05%). |
Warm-Up Time | Less than 3 minutes at room temperature. |
Operating Temperature | +18癈 to+28癈 (+65癋 to+82癋) for rated accuracy; the Model 828 may be operated from 癈 to +1 8癈 (+32癋 to +65癋) and +28癈 and +50癈 (+82癋 to +122癋) at derated specifications. |
Standard: | 115 V Operation: 104 to 127 volts, 47 to 63 Hz, 45 watts 230 V Operation: 207 to 253 volts, 47 to 63 Hz, 45 watts |
Optional: | 100 V Operation: 90 to 110 volts, 47 to 63 Hz, 45 watts 200 V Operation: 1 80 to 220 volts, 47 to 63 Hz, 45 watts |
Height: | 1 9 cm (71/2 inches) |
Width: | 21.6 cm (81/2 inches) |
Depth: | 33 cm (13 inches) |
Weight: | 25 lbs (11.3 kg) |
High burden capability is achieved by a rating of 10VA on the higher voltage ranges and a rating of 5VA on the one volt range. Compliance of the current ranges is 5V except for the 10A range where it is 1V. The percent error of a meter under test may be quickly determined and directly read by use of the Model 828's Percent Error control. Analog meters may be tested for "sticky" movements by use of the run-up mode which allows voltage or current to be smoothly and continuously applied to the meter under test. The calibrator supplies voltage and current at dc and at internally generated frequencies of 50, 60 and 400 Hz. Low ripple on dc functions and low distortion in ac modes minimizes random error when calibrating digital and rectifier type instruments. It is protected against accidental short circuit or overburdening of the output by internal sense circuitry which removes the output voltage or current if the load ratings are exceeded. High output stability makes this instrument suitable for use as a constant current or constant voltage source.
請輸入賬號
請輸入密碼
請輸驗證碼
以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,化工儀器網對此不承擔任何保證責任。
溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。