0081-C Mass Check ® Immunosuppressants whole blood control, four-level (I + II + III + IV)LC-MS/MS
- 公司名稱 北京祥生興業科技有限公司
- 品牌 CHROMSYSTEMS
- 型號 0081-C
- 產地
- 廠商性質 代理商
- 更新時間 2024/9/19 13:45:07
- 訪問次數 66
聯系方式:張先生13681184382 查看聯系方式
聯系我們時請說明是化工儀器網上看到的信息,謝謝!
溫馨提示:僅用于科研,不可用于臨床治療
- 英文名:
Mass Check ® Immunosuppressants whole blood control, four-level (I + II + III + IV)
- 庫存:
大量
- 供應商:
Chromsystems祥生興業
- 保存條件:
低溫保存
- 規格:
4 x 2 x 2 ml
- Chromsystems提供實驗室診斷試劑盒適用于HPLC和LC-MS/MS。.產品范圍包括完整的試劑盒和質控、校準。
MassCheck® Immunosuppressants Whole Blood Controls
Description
MassCheck® Immunosuppressants Whole Blood Controls, lyophilised.
For MassTox® Immunosuppressants in Whole Blood.
0081 Four-Level, 4 x 2 x 2 ml
0082 Level I, 5 x 2 ml
0083 Level II, 5 x 2 ml
0084 Level III, 5 x 2 ml
0085 Level IV, 5 x 2 ml
0089 Blank Control, 5 x 2 ml
These lyophilised quality controls from Chromsystems are based on human whole blood and should be used according to the instruction manual. They are designed to monitor the accuracy and precision of analytical procedures in clinical diagnostics for the quantitative determination of cyclosporin A, rapamycin, tacrolimus and everolimus in whole blood. While following the instructions the controls should be handled and measured in the same manner as a patient specimen.
Technical Data
Method of Analysis
Concentrations
Please note: Concentration values shown here are typical. For lot specific concentration values please refer to the product insert supplied with the materials. The product inserts can also be requested from us directly or downloaded from our website after login.
Level I | Level II | Level III | Level IV | |||||||
---|---|---|---|---|---|---|---|---|---|---|
Substance | Method | Unit | Value | Range | Value | Range | Value | Range | Value | Range |
Cyclosporin A | HPLC-UV | g/l | 96 | 77-115 | 242 | 193-290 | 476 | 381-572 | 1899 | 1519-2278 |
Dade-Behring EMIT | g/l | 101 | 81-121 | 247 | 198-296 | 467 | 374-560 | 2051 | 1641-2461 | |
Abbott TDx | g/l | 110 | 83-138 | 274 | 198-296 | 528 | 396-659 | 1833 | 1375-2291 | |
Rapamycin (Sirolimus) | LC-MS/MS | g/l | 3.7 | 2.6-4.8 | 9.1 | 7.3-10.9 | 18.0 | 14.4-21.6 | 39.2 | 31.4-47.0 |
Tacrolimus (FK 506) | Abbott IMx | g/l | 6.4 | 5.1-7.7 | 12.2 | 9.8-14.6 | 23.7 | 19.0-28.4 | 44.6 | 35.7-53.5 |
LC-MS/MS | g/l | 3.4 | 2.7-4.1 | 8.2 | 6.6-9.8 | 18.1 | 14.5-21.7 | 36.0 | 28.8-43.2 | |
Everolimus | LC-MS/MS | g/l | 2.5 | 1.6-3.3 | 4.4 | 2.9-6.0 | 8.9 | 5.7-11.9 | 30.1 | 21.1-39.1 |
Reconstitution
Carefully remove the metal seal and rubber stopper from the vials. Add exactly 2.0 ml of distilled water into each vial. Return the stopper and incubate for 10 to 15 minutes at room temperature. Swirl the vials to dissolve the contents until homogeneity.
Storage conditions
Stored below -18 °C and unopened, the lyophilised controls are stable until the expiry date specified on the vial labels. The reconstituted controls can be stored up to 1 week light protected and tightly capped at +2 to +8 C. If the controls cannot be used within this period, they should be aliquoted and stored frozen below -18 °C (up to 9 months). For more details about storage conditions please check your instruction manual.